The device is on.
• Press on "Backlight" to activate the device backlight.
• Press again on this key to deactivate the backlight.
Activate the backlight
• Remove the front part at the back of the device.
• Change the old batteries by AAA LR03 1.5V batteries.
• Replace the front.
Change the batteries
Adjust the coefcient
This coefcient corresponds to the coefcient of the differential pressure element used to measure the velocity (Pitot L
tube (coeff: 1.0015) or S (coeff: 0.84), Debimo blades...).
The device is on and displays the measurement.
• Press on "Select" until "COEF" blinks on screen.
• Press on "OK".
The rst digit blinks.
• Press on "Select" to select the rst digit value and press on "OK" to validate, performs the same procedure for the
following digits (press on "On/Off/Esc", to return to the previous digit).
• When the rst digit is adjusted, press on "OK" to validate.
"COEF" blinks on screen.
• Press on "On/Off/Esc" to return to the measurement.
Adjust the temperature
To have a precise and representative measurement, the temperature of the location where are performed the
measurements must be entered in the device. The temperature must be between -20 and +80 °C.
The device is on and displays the measurement.
• Press on "OK".
The rst digit blinks.
• Press on "Select" to select the temperature sign: - for a negative temperature and 0 for a positive temperature.
• Press on "OK" to validate and pass to the following digit.
• Press on "Select" to select the value of this second digit and press on "OK" to validate, perform the same procedure for
the following digit (press on "On/Off/Esc", to return to the previous digit).
• When the last digit is adjusted, press on "OK" to validate.
"TEMP" blinks on screen.
• Press on "On/Off/Esc" to return to the measurement display.
Adjust the atmospheric pressure
To have a precise and representative measurement, the atmospheric pressure of the location where are performed the
measurements must be entered in the device. The atmospheric pressure must be between 800 and 1200 hPa.
The device is on and displays the measurement.
• Press on "OK".
The rst digit of the atmospheric pressure blinks.
• Press on "Select" to select its value.
• Press on "OK" to validate and pass to the following digit, perform the same procedure for the following digits (press on
"On/Off/Esc"to return to the previous digit).
• When the rst digit is adjusted, press on "OK" to validate.
"P_ATM" blinks on screen.
• Press on "On/Off/Esc" to return to the measurement display.