sauermann CO 110 Kasutusjuhend

Kategooria
Mõõtmine, testimine
Tüüp
Kasutusjuhend
1
2
3
4
5
COmeter
USER MANUAL
Description of the device
1. OK key
2. Backlight key
3. Select key
4. On/Off/Esc key
5. Hold/min/max key
CO 110
Technical specications
Measuring element C0: electrochemical sensor. Temperature: NTC
Display
4 lines, LCD technology. Dimensions 50 x 36 mm
2 lines of 5 digits with 7 segments (value)
2 lines of 5 digits with 16 segments (unit)
Connector Retractable, 0.45 m length, extension: 2.4 m
Housing ABS, protection IP54
Keypad 5 keys
European directives 2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS II; 2012/19/EU WEEE
Power supply 4 batteries AAA LR03 1.5 V
Battery life 200 hours
Ambience Neutral Gas
Conditions of use (°C,%HR, m) From 0 to +50 °C. In non condensing conditions. From 0 to 2000 m.
Storage temperature From -20 to +80 °C
Auto shut-off Adjustable from 0 to 120 min
Weight 310 g
General features
Parameters Measuring units Accuracy(1) Measuring range Resolution
CO ppm ±3 ppm
±3% of reading
From 0 to 100 ppm
From 100 to 500 ppm 0.1 ppm
Temperature °C, °F ±0.4% of reading
±0.3 °C From -20 to +80 °C 0.1 °C
(1) All the accuracies indicated in this technical datasheet were stated in laboratory conditions, and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation.
Once the measurements are frozen:
Press “Hold/min/max”.
The device displays the maximum value of CO measured since the last questioning at the top of the screen and the
minimum value of CO measured since the last questioning at the bottom of the screen.
Press again “Hold/min/max”.
The device displays the maximum value of temperature measured since the last questioning at the top of the screen and
the minimum value of temperature measured since the last questioning at the bottom of the screen.
Press “On/Off/Esc” to return to the measurements display.
Display the minimum and the maximum
Turn on the device by pressing on “On/Off/Esc”.
The device displays its name “CO 110” then performs an autozero during 5 seconds.
The probe must be left to the open-air during this autozero.
Put the probe to the required location.
The device displays the CO and temperature measurements.
Perform a measurement
During the measurement:
Press “Hold/min/max” to freeze the measurement.
“Hold” appears on the screen and the measurements are frozen.
Press “On/Off/Esc” to exit the hold function.
The device returns to the measurements display.
Freeze the measurement
Adjust the device
Device drop-down menu
During the measurement:
Press “OK”.
“CO MAX” displays on screen with below the measuring time to adjust. This time is between 0:00 and 5:00.
Press “Select” to adjust the rst digit then “OK” to validate, perform the same procedure for the following digits.
When the st digit is adjusted, press “OK”.
“START” displays on screen.
Press again “OK” to start the measuring time.
The device displays the measuring time count. At the end of this count, the CO max measurement displays on screen.
Perform a CO max measurement
Adjust the auto shut-off
The device is on and displays the measurements.
Press “Select” until AUTO OFF” blinks on screen.
Press “OK”.
The time before device auto shut-off blinks on screen.
Press “Select” to select the time before device auto shut-off: 15, 30, 45, 60, 75, 90, 105, 120 or OFF.
Press “OK” to validate.
AUTO OFF” blinks on screen.
Press “On/Off/Esc” to return to the measurements display.
Activate or deactivate the keys beep
The device is on and displays the measurements.
Press “Select” until “BEEP” blinks on the screen.
Press “OK”.
“ON” or “OFF” blinks on the screen.
Press “Select” to activate the keys beep “ON” or deactivate the keys beep “OFF”.
Press “OK”.
“BEEP” blinks on the screen.
Press “On/Off/Esc” to return to the measurement display.
Select the measuring unit in temperature
Press “Select”.
“UNIT” blinks on screen.
Press “OK”.
The unit currently used blinks at the bottom of the screen.
Press “Select” until the required unit appears: °C or °F.
Press “OK” to validate.
“UNIT” blinks on screen.
Press “On/Off/Esc” to return to the measurements display.
NT_EN – CO110 – 10/11/17– Non-contractual document – We reserve the right to modify the characteristics of our products without prior notice.
www.sauermanngroup.com
The device is on and displays the measurements.
Press “Backlight” to activate the device backlight.
To deactivate it press again “Backlight”.
Activate the backlight
Remove the front part at the back of the device.
Change the old batteries by AAA LR03 1.5.V batteries.
Replace the front part.
Change the batteries
Adjust the alarms
The device is on and displays the measurements.
It’s possible to adjust 1 or 2 alarms which will be triggered when the alarm threshold has been reached and/or exceeded.
This threshold is between 0 and 500 ppm.
Press “Select” until AL 1” blinks on screen. AL 1” corresponds to the level 1 alarm.
Press “OK”.
The rst alarm digit blinks.
Press “Select” to select the rst digit value: between 0 and 5 or OFF to deactivate the alarm.
Press “OK” to validate and to pass to the following digit, perform the same procedure for the following digits (press
“On/Off/Esc” to return to the previous digit).
Press “OK” when the last digit is adjusted.
AL 1” blinks on screen.
Press “On/Off/Esc” to return to the measurement display.
To adjust the level 2 alarm, press “Select” until AL 2” blinks on screen then follow the same procedure as the
adjustment of the level 1 alarm.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

sauermann CO 110 Kasutusjuhend

Kategooria
Mõõtmine, testimine
Tüüp
Kasutusjuhend

teistes keeltes