Pt100 thermometers
USER MANUAL
Device description
1. Backlight key
2. OK key
3. On/Off/Esc key
4. Selection key
5. Hold/min/max key
TR 61 – TR 62
Technical specications
General features
Measuring element Pt100 class A
Number of channels TR61: 1 channel / TR62: 2 channels
Display 4 lines, LCD technology. Size: 50 x 36 mm.
2 lines of 5 digits with 7 segments (value) / 2 lines of 5 digits with 16 segments (unit)
Housing ABS, protection IP54
Keypad 5 keys
Connector Compensated miniature female connectors
European directives 2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS II; 2012/19/EU WEEE
Power supply 4 batteries AAA LR03 1.5 V
Battery life 180 hours
Ambience Neutral gas
Conditions of use (°C, %RH, m) From 0 to +50 °C. In non condensing conditions. From 0 to 2000 m.
Storage temperature From -20 to +80 °C
Auto shut-off Adjustable from 0 to 120 min
Weight 210 g
(1) All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with required compensation.
Pt100 probes (see related data sheet)
Measuring units Measuring range Accuracy1Resolution
°C, °F From -100 to +400 °C ±0.4% of reading ±0.3 °C 0.1 °C
During a measurement:
Press “Hold/min/max”.
“Hold” is displayed and the measurement is frozen.
Press again “Hold/min/max”.
The device displays the maximum value measured since the last questioning of the frozen channel at the top of the
screen and the minimum value measured since the last questioning of the frozen channel at the bottom of the screen.
For a TR 62 with two connected probes:
The device displays the maximum value of the C1 probe measured since the last questioning of the frozen channel at
the top of the screen and the minimum value of the C1 probe measured since the last questioning of the frozen channel
at the bottom of the screen.
Press again ”Hold/min/max”.
The device displays the maximum value of the C2 probe measured since the last questioning of the frozen channel at
the top of the screen and the minimum value of the C2 probe measured since the last questioning of the frozen channel
at the bottom of the screen.
Press again Hold/min/max.
The device displays the maximum value of the delta T since the last questioning of the frozen channel at the top of the
screen and the minimum value of the delta T since the last questioning of the frozen channel at the bottom of the screen.
Press “On/Off/Esc” to return to the measurement display.
Turn the device on by pressing “On/Off/Esc”.
Put the probe(s) to the required location.
The device displays the temperature measurement.
If the thermometer used is the TR62 with two connected probes, the device will display the two measurements.
Press “OK“ to display the delta T between the two measured temperatures.
Perform a meaurement
Hold the measurement, display the maximum and minimum
Select the temperature unit
The device is on and displays the measurement.
Press “Selection” until “UNIT” blinks on screen. Press “OK”.
The unit currently used blinks on screen.
Press “Selection” until the required unit appears: °C or °F.
Press “OK” to conrm the unit selection.
“UNIT” blinks on screen.
Press ”On/Off/Esc” to return to the measurement display.
Set the auto shut-off
The device is on and displays the measurement.
Press “Selection” until AUTO OFF” blinks on screen. Press “OK”.
The time before the device auto shut-off blinks at the bottom of the screen.
Press “Selection” until the required time before the device auto shut-off appears: 15, 30, 45, 60, 75, 90, 105, 120
minutes or OFF.
Press “OK” to conrm the selection.
AUTO OFF” blinks on screen.
Press ”On/Off/Esc” to return to the measurement display.
Activate or deactivate the keys beep
The device is on and displays the measurement.
Press “Selection” until “BEEP” blinks on screen. Press “OK”.
“OFF” or “ON” blinks on screen.
Press “Selection” to activate the keys beep “ON” or deactivate it “OFF”.
Press “OK” to conrm the selection.
“BEEP” blinks on screen.
Press ”On/Off/Esc” to return to the measurement display.
Set the device
Set the alarm
This function allows to activate a temperature alarm on channel 1 when the device reaches or exceeds a previously
dened threshold. This threshold is between -100 and +400 °C.
The device is on and displays the measurement.
Press “Selection” until AL 1” blinks on screen.
Press “OK”.
The rst digit blinks. This digit also allows to deactivate the alarm.
Press “Selection” to select the value of the rst digit: -, 0, 1 or OFF.
If “Off is selected, the alarm is deactivated. Press “OK” to conrm, AL 1” blinks on screen.
If -, 0 or 1 is selected, press “OK”. The next digit blinks, press “Selection” to select its value then press “OK” to conrm.
Perform the same procedure to select the value of the following digits (press “On/Off/Esc” to return to the previous
digit). Press “OK” when the last digit is set, AL 1” blinks on screen.
Press ”On/Off/Esc” to return to the measurement display.
The device is on.
Press “Backlight” to activate the device backlight.
Press again this key to deactivate the backlight.
Activate the backlight
Remove the front part at the back of the device.
Change the old batteries by AAA LR03 1.5V batteries.
Replace the front
Replace the batteries
To turn off the device:
Carry out a long press on “On/Off/Esc”.
Turn off the device