sauermann LV 110 Kasutusjuhend

Tüüp
Kasutusjuhend
1
2
3
4
5
Vane probe thermo-anemometer
USER MANUAL
Description of the device
1. OK key
2. Backlight key
3. Select key
4. Backlight key On/Off/Esc key
5. Hold/min/max key
LV 110 - LV 111- LV 117
Technical specications
General features
Measuring units
LV 111 - LV 117 - LV 110: m/s, fpm, km/h
All models (airow): m³/h, cfm, l/s, m³/s
All models (temperature): °C, °F
Measuring elements Air velocity: Hall effect sensor / Ambient temperature: NTC sensor
Display
4 lines, LCD technology. Dimensions 50 x 36 mm.
2 lines of 5 digits with 7 segments (value)
2 lines of 5 digits with 16 segments (unit)
Vane diameter LV111: Ø 14 mm / LV117: Ø 70 mm / LV110: Ø 100 mm
Cable Coiled, 0.45 m length, extension 2.4 m
Housing ABS, protection IP54
Keypad 5 keys
European directives 2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS II; 2012/19/EU WEEE
Power supply 4 batteries AAA LR03 1.5 V
Battery life 58 hours(1)
Ambience Neutral gas
Conditions of use (°C, %RH, m) From 0 to +50 °C. In non condensing conditions. From 0 to 2000 m.
Operating temperature (probe) From 0 to +50 °C
Storage temperature From -20 to +80 °C
Auto shut-off Adjustable from 0 to 120 min
Weight 390 g
Parameters Accuracy(1) Measuring range Resolution
Air velocity
LV111 (Ø 14 mm): from 0.8 to 3 m/s: ±3% of reading ±0.1 m/s
From 3.1 to 25 m/s: ±1% of reading ±0.3 m/s
LV110 (Ø 100 mm): from 0.3 to 3 m/s: ±3% of reading ±0.1 m/s
From 3.1 to 35 m/s: ±1% of reading ±0.3 m/s
LV 117 (Ø 70 mm): from 0.4 to 3 m/s: ±3% of reading ±0.1 m/s
From 3.1 to 35 m/s: ±1% of reading ±0.3 m/s
From 0.8 to 25 m/s
From 0.3 to 35 m/s
From 0.4 to 35 m/s
0.1 m/s
0.01 m/s, 0.1 m/s
0.1 m/s
Airow
All models ±3 % of reading ±0.03 x area (cm²) From 0 to 99 999 m³/h 1 m³/h
Temperature
All models ±0.4 % of reading ±0.3 °C From -20 to +80 °C 0.1 °C
(1)All the accuracies indicated in this technical datasheet were stated in laboratory conditions, and can be guaranteed for measurements carried out in the same conditions, or carried out with calibration compensation.
Once the measurement is frozen:
Press on "Hold/min/max".
The device displays the maximum value of velocity or airow measured since the last questioning at the top of the
screen and the minimum value of velocity or airow measured since the last questioning at the bottom of the screen.
Press again on "Hold/min/max".
The device displays the maximum value of temperature measured since the last questioning at the top of the screen and
the minimum value of temperature measured since the last questioning at the bottom of the screen.
Press on "On/Off/Esc" to return to the measurements display.
Display the minimum and the maximum
Turn on the de device by pressing on "On/Off/Esc".
The device displays its name "LV110", "LV111" or "LV117".
Put the probe to the required location.
The device displays the velocity and temperature measurement.
Perform a measurement
During a measurement:
Press once on "Hold/min/max".
"Hold" appears on screen and the measurement is frozen.
Press on "On/Off/Esc" to exit the hold function.
The device returns to the measurement display.
Freeze the measurement
The device is on and displays the measurement.
Press on "OK".
"START" displays on screen.
Press on "OK", the average calculation is launched, "END" displays on screen,"AVG" blinks.
Press on"OK" to stop the calculation, the average displays.
Press on "OK" to display the maximum and minimum values measured during the calculation of this average.
Press on "On/Off/Esc" to exit the average function and return to the measurements display.
Perform an average in velocity
It is possible to display and to measure the airow in place of the velocity.
The device is on and displays the velocity and temperature measurements.
Press on "Select".
"MODE" blinks on screen.
Press on "OK".
"VELO" blinks at the bottom of the screen.
Press on "Select"to select the "FLOW" mode then press on "OK" to validate.
Depending on the device type, "CIRC", "K 25 85", "CONE" or "RECT" blinks at the bottom of the screen.
Press on "Select" to select rectangular "RECT", "CIRC", "K 25 85" (LV110) or "CONE" (LV111) for a measurement
with an airow cone then press on"OK".
If "RECT" or "CIRC" is selected, the measuring unit of the type of duct displays.
Press on "Select" to select the unit: mm or inch then press on "OK" to validate.
If the type of selected duct is circular: "DIAM" displays and the rst digit blinks.
Press on "Select" to select its value then press on "OK" to validate. Perform the same procedure to select the values
of the following digits (to return to the previous digit press on "On/Off/Esc"). When the last digit is adjusted, press on
"OK" to validate. The device returns to the display of airow and temperature measurements.
If the type of selected duct is rectangular: "LENGT" displays on screen and the rst digit blinks.
Press on "Select" to select its value then press on "OK" to validate. Perform the same procedure to select the values of
following digits (to return to the previous digit press on "On/Off/Esc"). When the last digit is adjusted, press on "OK".
Display and measure the airow
Adjust the device
Device drop-down menu
Select the measuring unit in velocity
The device is on and in velocity mode.
Press on "Select" until "UNIT" blinks on screen.
Press on "OK".
The unit currently used blinks on screen.
Press on "Select" until the required unit appears: m/s, fpm or km/h.
Press on "OK" to validate the unit selection.
"UNIT" blinks on screen.
Press on "On/Off/Esc" to return to the measurements display.
"WIDTH" to adjust the width displays on screen, perform the same procedure to select the digits value, when the last digit
is adjusted, press on "OK" to validate. The device returns to the display of airow and temperature measurements.
If "K 25 85" is selected (LV110): the device returns to the measurements display.
If "CONE" is selected (LV111): "CONE" displays on screen and the cone type blinks: K35, K75, K120, K150.
Press on "Select" to select the type of cone, then press on "OK" to validate. The device returns to the display of airow
and temperature measurements.
The length, the width and the section diameter are adjustable from 1 to 3000 mm.
Select the measuring unit in airow
The device is on and in airow mode.
Press on "Select" until "UNIT" blinks on screen.
Press on "OK"
The unit used currently blinks on screen.
Press on "Select" until the required unit appears: m3/h, L/s, cfm or m3/s.
Press on "OK" to validate the unit selection.
"UNIT" blinks on screen.
Press on "On/Off/Esc" to return to the measurements display.
NT_EN – portable-LV110-LV111-LV117 – 10/11/17 – Non-contractual document – We reserve the right to modify the characteristics of our products without prior notice.
www.sauermanngroup.com
The device is on.
Press on "Backlight" to activate the device backlight.
Press again on this key to deactivate the backlight.
Activate the backlight
Remove the front part at the back of the device.
Change the old batteries by AAA LR03 1.5V batteries.
Replace the front part.
Change the batteries
Select the measuring unit in temperature
The device is on and displays the measurements.
Press on "Select" until "UNIT" blinks on screen with the unit temperature displayed below.
Press on "OK".
The unit currently used blinks at the bottom of the screen.
Press on "Select" until the required unit appears: °C or °F.
Press on "OK" to validate.
"UNIT" blinks on screen.
Press on "On/Off/Esc" to return to the measurements display.
Adjust the auto shut-off
The device is on and displays the measurement.
Press on "Select" until "AUTO OFF" blinks on screen.
Press on "OK".
The time before device auto shut-off blinks at the bottom of the screen.
Press on "Select" until the required time before auto shut-off appears: 15, 30, 45, 60, 75, 90, 105, 120 minutes or OFF
Press on "OK" to validate.
"AUTO OFF" blinks on screen.
Press on "On/Off/Esc" to return to the measurement.
Activate or deactivate the keys beep
The device is on and displays the measurement.
Press on "Select" until "BEEP" blinks on screen.
Press on "OK".
"OFF" or "ON" blinks on screen.
Press on "Select" to activate the keys beep "ON" or deactivate it "OFF".
Press on "OK" to validate.
"BEEP" blinks on screen.
Press on "On/Off/Esc" to return to the measurement display.
  • Page 1 1