Phcbi MPR-1014R Kasutusjuhend

Tüüp
Kasutusjuhend

See juhend sobib ka

Please read the operating instructions carefully before using this product, and keep the operating
instructions for future use.
See page 44 for all model numbers.
Operating Instructions
Pharmaceutical Refrigerator
MPR-1014R
1
CONTENTS
INTRODUCTION ································································2
PRECAUTIONS FOR SAFE OPERATION ································3
ENVIRONMENTAL CONDITIONS ··········································7
REFRIGERATOR COMPONENTS
Refrigerator unit ·····························································8
Control panel components ·············································· 10
INSTALLATION SITE ························································ 11
INSTALLATION ································································ 12
START-UP OF UNIT ························································· 14
CHAMBER TEMPERATURE SETTING ································· 15
LOCK OF CHAMBER TEMPERATURE ································· 16
DEFROST OF EVAPORATOR ············································ 17
ALARM TEMPERATURE SETTING
Setting of high temperature alarm ···································· 18
Setting of low temperature alarm ······································ 19
SETTING OF DELAY OF DOOR ALARM ······························· 20
SETTING OF RINGBACK OF ALARM BUZZER ······················ 21
OPERATION CHECK AFTER RECOVERY ···························· 21
REMOTE ALARM TERMINAL ············································· 22
ALARM FUNCTIONS ························································· 23
Air circulation alarm ······················································ 23
SAFETY FUNCTIONS ······················································· 23
SELF DIAGNOSTIC FUNCTIONS ········································ 24
ROUTINE MAINTENANCE
Cleaning of exterior, interior, and accessories ····················· 25
Replacement of fluorescent lamp ····································· 26
Replacement of glow starter ············································ 26
Cleaning of evaporating tray ··········································· 27
Cleaning of filter ··························································· 28
TROUBLESHOOTING ······················································· 29
DISPOSAL OF UNIT ························································· 30
Decontamination of unit ················································· 30
TEMPERATURE RECORDER ············································· 35
INSTALLATION OF TEMPERATURE RECORDER
Attachment of MTR-0621LH ············································ 36
Attachment of MTR-G04A or MTR-G04C ··························· 38
BATTERY FOR POWER FAILURE ALARM ···························· 40
SETTING FOR BATTERY BEFORE STARTING ······················ 42
SPECIFICATIONS ···························································· 43
PERFORMANCE ······························································ 44
SAFETY CHECK SHEET···················································· 45
2
INTRODUCTION
Read the operating instructions carefully before using the appliance and follow the instructions for
safety operation.
PHC Corporation never guarantees any safety if the appliance is used for any objects other than
intended use or used by any procedures other than those mentioned in the operating instructions.
Keep the operating instructions in an adequate place to refer to it as necessary.
The contents of the operating instructions will be subjected to change without notice due to the
improvement of performance or functions.
Contact our sales representative or agent if any page of the operating instructions is lost or page order
is incorrect.
Contact our sales representative or agent if any point in the operating instructions is unclear or if there
are any inaccuracies.
No part of the operating instructions may be reproduced in any form without the expressed written
permission of PHC Corporation.
IMPORTANT NOTICE
PHC Corporation guarantees this product under certain warranty conditions. However, please note
that PHC Corporation shall not be responsible for any loss or damage to the contents of the product.
<Intended Use>
This equipment is designed for storage of pharmaceuticals, samples and reagents.
3
PRECAUTIONS FOR SAFE OPERATION
It is imperative that the user complies with the operating instructions
as it contains important safety advice.
Items and procedures are described so that you can use this unit correctly and safely.
If the precautions advised are followed, this will prevent possible injury to the user and
any other person.
Precautions are illustrated in the following way:
WARNING
Failure to observe WARNING signs could result in a hazard to personnel
possibly resulting in serious injury or death.
CAUTION
Failure to observe CAUTION signs could result in injury to personnel and
damage to the unit and associated property.
Symbol shows;
This symbol means caution.
This symbol means an action is prohibited.
This symbol means an instruction must be followed.
Be sure to keep the operating instructions in a place accessible to users of this unit.
< Label on the unit >
This mark is labeled on the cover in which the electrical components of high voltage
are enclosed to prevent the electric shock.
The cover should be removed by a qualified engineer or service personnel only.
USA Only: This product has a fluorescent lamp that contains mercury. Disposal may be regulated in your
community due to environmental considerations. For disposal or information, please visit PHC website:
https://www.phchd.com.
Contains mercury / Contenu avec mercure
For more information on safe handling
procedures, the measures to be taken in
case of accidental breakage and safe
disposal options visit:
ec.gc.ca/mercure-mercury/.
Dispose of or recycle in accordance with
applicable laws.
Pour plus de renseignements sur les
procédures de manutention sécuritaire, les
mesures à prendre en cas de bris accidentel et
les options d’élimination sécuritaire visitez:
ec.gc.ca/mercure-mercury/.
Mettez au rebut ou recyclez conformément
aux lois applicables.
4
PRECAUTIONS FOR SAFE OPERATION
Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to
rain water.
Only qualified engineers or service personnel should install the unit. The installation by
unqualified personnel may cause electric shock or fire.
Install the unit on a sturdy floor and take an adequate precaution to prevent the unit from
turning over. If the floor is not strong enough or the installation site is not adequate, this may result
in injury from the unit falling or tipping over.
Never install the unit in a humid place or a place where it is likely to be splashed by water.
Deterioration of the insulation may result which could cause current leakage or electric shock.
Never install the unit in a flammable or volatile location. This may cause explosion or fire.
Never install the unit where acid or corrosive gases are present as current leakage or electric
shock may result due to corrosion.
Always ground (earth) the unit to prevent electric shock. If the power supply outlet is not
grounded, it will be necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
Connect the unit to a power source as indicated on the rating label attached to the unit. Use
of any other voltage or frequency other than that on the rating label may cause fire or electric shock.
Never store volatile or flammable substances in this unit if the container cannot be sealed. These
may cause explosion or fire.
Do not insert metal objects such as a pin or a wire into any vent, gap or any outlet on the unit.
This may cause electric shock or injury by accidental contact with moving parts.
Use this unit in safe area when treating the poison, harmful or radiate articles. Improper use
may cause bad effect on your health or environment.
Turn off the power switch (if provided) and disconnect the power supply to the unit prior to any
repair or maintenance of the unit in order to prevent electric shock or injury.
Do not touch any electrical parts (such as power supply plug) or operate switches with a wet
hand. This may cause electric shock.
WARNING
5
PRECAUTIONS FOR SAFE OPERATION
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
Never splash water directly onto the unit as this may cause electric shock or short circuit.
Never put containers with liquid on the unit as this may cause electric shock or short circuit when
the liquid is spilled.
Never bind, process, or step on the power supply cord, or never damage or break the power
supply plug. A broken power supply cord or plug may cause fire or electric shock.
Do not use the power supply cord if its plug is loose. Such power supply cord may cause fire or
electric shock.
Never disassemble, repair, or modify the unit yourself. Any such work carried out by an
unauthorized person may result in fire, or electric shock or injury due to a malfunction.
Disconnect the power supply plug if there is something wrong with the unit. Continued
abnormal operation may cause electric shock or fire.
When removing the power supply plug from the power supply outlet, grip the power supply
plug, not the cord. Pulling the power supply cord may result in electric shock or fire by short circuit.
Disconnect the power supply plug before moving the unit. Take care not to damage the power
supply cord. A damaged power supply cord may cause electric shock or fire.
Disconnect the power supply plug when the unit is not used for long periods. Keeping the
connection may cause electric shock, current leakage, or fire due to the deterioration of insulation.
If the unit is to be stored unused in an unsupervised area for an extended period, ensure that
children do not have access and that doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel. Remove doors to
prevent accidents such as suffocation.
Do not put the packing plastic bag within reach of children as suffocation may result.
Do not position this unit and the other unit so that it is difficult to operate the disconnection of
the power supply plug. Failure to disconnect the power supply plug may cause fire if there is
something wrong with the unit.
WARNING
6
PRECAUTIONS FOR SAFE OPERATION
This unit must be plugged into a dedicated circuit protected by branch circuit breaker.
Use a dedicated power source as indicated on the rating label attached to the unit. A
multiple-tap may cause fire resulting from abnormal heating.
Connect the power supply plug to the power source firmly after removing the dust on the plug.
A dusty power supply plug or improper insertion may cause a heat or ignition.
Never store corrosive substances such as acids or alkalis in the unit except in a sealed
container. These may be detrimental to health and may cause corrosion of internal components,
cooling circuit or electrical components.
Check the setting when starting up of operation after power failure or turning off of power
switch. The stored items may be damaged due to the change of setting.
Be careful not to tip over the unit during movement to prevent damage or injury.
Prepare a safety check sheet when you request any repair or maintenance for the safety of service
personnel.
CAUTION
7
ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe at least under the following conditions (based on the IEC 61010-1):
Indoor use;
Altitude up to 2000 m;
Temperature 5 oC to 40 oC
Maximum relative humidity 80 % for temperature up to 31 oC decreasing linearly to 50 % relative
humidity at 40 oC;
Mains supply voltage fluctuations up to ±10 % of the nominal voltage;
Transient overvoltages up to the levels of OVERVOLTAGE CATEGORY II;
Temporary OVERVOLTAGES occurring on the mains supply;
Applicable pollution degree of the intended environment (POLLUTION DEGREE 2 in most cases);
8
REFRIGERATOR COMPONENTS
Refrigerator unit
14
Circuit breaker
switch
15
Remote alarm
terminal
[page 22]
10
1
MPR-1014R
5
6
4
7
9
11
12
3
2 13
3
8
Light switch
Space for temperature
recorder
[page 35]
Glow starter
[page 26] Fluorescent
lamp
[page 26]
9
REFRIGERATOR COMPONENTS
1. Control panel: The chamber temperature and alarms can be set through the keys on the control
panel. The temperature display and indicators are also provided on the control panel. Refer to page 10
for details.
2. Location of battery kit for power failure alarm: The battery kit (optional) alerts the power failure by
the alarm indicator and alarm buzzer.
3. Door switch: This switch detects the door status (open/close). The door check lamp is ON when
the door is open.
4. Door: Sliding type. The recessed portion on the rail enables the self-closing of the door. The glass
is pair construction.
The condensation may be found on the door frame with ambient humidity of about 50%R.H. and on the
glass surface with ambient humidity of about 60%R.H. This is not a malfunction.
The condensation on the door is dropped on the rail and exhausted to the evaporating tray.
5. Air exhaust vent: Do not block this vent. Arrange the stored items not to subject to the cold air
from the vent.
6. Drawer (Right side of MPR-1014R): To pull out the drawer, take out the drawer with pushing the
button at the bottom of the drawer.
<Important> Never pull out the multiple drawers at a time. It may cause tipping over of the unit.
7. Air intake vent (front bottom): Do not block this vent. Blocking this vent may cause unstable
chamber temperature. Do not insert fingers or similar articles into the vent.
8. Filter: Clean the filter once a month. Refer to page 28.
9. Evaporating tray: Defrost water from the evaporator accumulates in the tray and evaporates into the
atmosphere. See page 27 for cleaning.
10. Leveling foot (front, 3 locations): These are used for install the unit. Adjust the height of the
leveling feet by turning the screw bolts until 3 front casters are away from the floor.
11. Shelf: The set location is adjustable.
Items to be stored in the chamber must be placed on the shelves. Do not put stored items directly on
the bottom of the chamber.
12. Access port: This port allows a sensor or cable of measuring equipment to enter the chamber from
outside.
Replace insulation and the rubber caps when the access port is not used. Improper replacement may
cause rise of chamber temperature or condensation around the access port.
13. Exclusive alarm sensor: A sensor for detecting the temperature rise at the upper area of the
chamber. Refer to page 23 for alarm functions.
14. Back spacer (also used as a fixture): To keep the space between the unit and the back wall for
adequate cooling performance. And this can be used as a fixture. Fix the unit by using the fixture and
a rope or chain. Refer to page 12.
15. Circuit breaker switch: Switch ON this circuit breaker switch before the unit starts to run. Refer to
page 13.
10
REFRIGERATOR COMPONENTS
Control panel components
1. Door check indicator (DOOR): The red
LED lamp is lit when the door is opened.
2 minutes after the door check indicator is lit,
the buzzer is activated to notice the door opening.
[page 20]
2. Alarm indicator (ALARM): The red LED
lamp blinks during an alarm condition. [page 23]
3. Temperature display: Normally shows the
present chamber temperature and during an
alarm condition, shows an error code. [page 24]
4. Up arrow key ( ):
At “temperature display mode”; pressing this key
for more than 5 seconds leads setting mode.
At “setting mode”; increases number values
between choices. [page 15 to page 21]
At “setting of lock function”; selects ON-OFF of
key lock for chamber temperature setting. [page
16]
5. Scroll key ( ):
At “temperature display mode”; pressing this key
for more than 5 seconds leads “setting of lock
function”. [page 16]
At “setting mode”; moves the operator through
digit on the display. [page 15 to page 21]
6. Set key (SET):
At “temperature display mode”; pressing this key
leads “chamber temperature setting mode”.
At “setting mode”; stores the value into computer
memory.
7. Alarm buzzer stop key (BUZZER): Mutes
the audible alarm however, the audible alarm
recovers after certain period (ringback time) if the
same alarm status continues. [page 21]
Pressing this key during the power failure displays
the chamber temperature for 5 seconds (with
MPR-48B1 installed).
“temperature display mode”: the status which
the temperature display shows the current
chamber temperature
“setting mode”; the status which the
temperature display is ready to be entered after
pressing the up arrow key for 5 seconds.
7 6 5 4
3 2 1
DOOR ALARM TEMPERATURE(oC)
BUZZE
R
SET
11
INSTALLATION SITE
To operate this unit properly and to obtain maximum performance, install the unit in a location with the
following conditions:
A location not compliance with the following conditions may cause poor performance, failure or
accident.
A location not subjected to direct sunlight
Do not install the unit under direct sunlight. Installation in a location subjected to direct sunlight may
degrade the performance.
A location with adequate ventilation
Leave at least 10 cm around the unit for ventilation. Poor ventilation will result in a reduction of the
performance and consequently the failure.
A location away from heat generating sources
Avoid installing the unit near heat-emitting appliances such as a heater or a boiler etc. Heat can
decrease the intended performance of the unit.
A location with little temperature change
Install the unit under stable ambient temperature. Installing the unit under unstable ambient temperature
may result in unstable performance.
A location with a sturdy and level floor
Always install the unit on a sturdy and level floor withstanding the total weight of the unit. The uneven
floor or tilted installation may cause failure or injury. Unstable condition may cause vibration or noise.
A location not prone to high humidity
Install the unit in the ambient of 80 % R.H. or less humidity. Installation under high humidity may cause
current leakage or electric shock.
CAUTION
The unit may collect excessive frost on the evaporator if it is installed in high temperature and high
humidity location. This will cause frequent defrosting.
A location without flammable or corrosive gas
Never install the unit in a flammable or corrosive location. This may cause explosion or fire or may result
in the current leakage or electric shock by the corrosion of the electrical components.
CAUTION
Never install the unit in a location where the corrosive material such as sulfur compound is likely
to be generated (near a drainage facilities, etc). Corrosion of the copper pipe may result in
deterioration of cooling unit and consequently the failure.
A location without the possibility of anything fall
Avoid installing the unit in the location where anything can fall down onto the unit. This may cause the
breakdown or failure of the unit.
12
INSTALLATION
1. After unpackaging
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted neutral dishwashing detergent.
Undiluted detergent can damage the plastic components. For the dilution, refer to the instruction of
the detergent.
After the cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the
panels with a dry cloth.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
2. Installing the unit
Extend the leveling feet by rotating them counterclockwise
to separate 3 front casters from the floor. [Fig. 1]
And adjust the leveling foot to level the unit.
The unit is installed firmly by separating the casters from
the floor. The unit may be moved at the time of
opening/closing the door if the casters contact with floor.
3. Fixing the unit
Fix the unit to the wall by using 2 back spacers [Fig. 2] and
a rope or chain.
The back spacers (which are also used to prevent the
unit from toppling over) have a dual purpose: To provide a
clearance between the unit and the wall behind it, and to
prevent the unit toppling over.
4. Ground (earth) the unit
The ground (earth) is for preventing the electric shock in
the case of the electrical insulation is somehow degraded.
Always ground the unit at the time of installation.
A 3-prong plug with grounding pole is provided to the
power supply cord. There is no need for electric work for
grounding.
If the power supply outlet is not grounded, it will be
necessary to install a ground by qualified engineers.
Leveling foot Fig. 1
Back spacer
Fig. 2
13
INSTALLATION
5. Installation of shelves
The shelves are packaged all together. Set the shelves in the standard position before start-up of the
unit.
Set the shelf stopper to the following location (5 locations) on the shelf support to adjust the shelves in the
standard position:
1) The sixth hole under the first screw from the top (top screw) [Fig. 3]
2) The fifth hole under the third screw from the top.
3) The fourth hole under the fifth screw from the top
4) The third hole under the third screw from the bottom.
5) The second hole under the first screw from the bottom (bottom screw).
<Important>
Never put the stored item on the bottom of the chamber. Always put them on the shelf or in the drawer
(in the case of MPR-1014R).
6. Turn on the circuit breaker switch
This unit is equipped with a circuit breaker on the back.
Make sure to turn on the circuit breaker switch before the
unit starts to run [Fig. 4].
When the operation of the unit is stopped by this breaker, contact our sales representative or agent after
disconnected the power supply plug.
Fi
g
. 3
Shelf support
Shelf stopper
Top screw
Circuit breaker
switch
Fi
g
. 4
14
START-UP OF UNIT
Follow the procedures for the initial and consequent operations of the unit.
At the recovery after power failure, the operation is start-up automatically with the setting before power
failure. [page 21]
1. Connect the power supply cord to the dedicated outlet with appropriate rating.
The alarm indicator and the chamber temperature may blink. (This is not a malfunction.)
<Important>
If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes.
This protects the compressor.
2. On start-up, the alarm buzzer sometimes operates. In this case, stop the alarm buzzer by pressing
the alarm buzzer stop key (BUZZER). (This is not a malfunction.)
<Important>
Close the door securely. The door check indicator is lit when the door is open. The alarm buzzer
activates when the door is opened for more than 2 minutes. The alarm buzzer is canceled when the
door is closed.
3. Allow the chamber temperature to fall to 5 oC (The chamber temperature is set to 5 oC at the factory).
Check the chamber temperature on the temperature display.
Allow the chamber temperature to fall to the set temperature when the start-up after cleaning,
maintenance or moving.
4. Switch on the light switch to check the fluorescent lamp is on. After checking, turn off the fluorescent
lamp if the light is not necessary.
5. Set the desired chamber temperature. [page 15]
6. When the chamber temperature gets to the set temperature, begin slowly placing items into the
chamber to minimize the temperature rise.
<Important>
Always put the stored items on the shelf or in the drawer, not on the bottom of the chamber. Do not
contact the stored items to the chamber walls.
Do not block the air intake vent nor air exhaust vent. Place the stored items with adequate space
between them to keep the cool air circulation.
15
CHAMBER TEMPERATURE SETTING
Set the chamber temperature according to the condition of use. This refrigerator can keep the stored
items for long period under appropriate temperature.
Setting range of chamber temperature: between 2 oC and 14 oC
Initial setting (factory setting): 5 oC
<Important>
The chamber temperature of 2 oC may cause partial freeze of stored items.
Example: Change the chamber temperature to 4 oC from 5 oC
Following shows a sample setting. Set the desired chamber temperature according to the condition
of use.
Description of operation Key operated Indication after operation
1
Connect the power supply plug to
the dedicated outlet.
(only when start-up of unit)
----- The current chamber temperature is
displayed.
2 Press set key. SET The current setting (005) is displayed
and the first digit blinks.
3 Press up arrow key and scroll the
figure to 4. The display is changed to 004 from
005.
4 Press set key. SET
Set temperature is memorized and
the current chamber temperature is
displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.
CAUTION
It is recommended to install an automatic temperature recorder (optional component) to check the
highest/lowest chamber temperature when the item which needs severe temperature control is stored.
16
LOCK OF CHAMBER TEMPERATURE
The setting of chamber temperature can be protected to avoid an accidental change. When the lock is
ON, the change of chamber temperature setting through the key pad is not available.
Initial setting (factory setting): lock OFF
Display Mode Function
L 0 Lock is OFF Enable to change the chamber temperature setting
L 1 Lock is ON Disable to change the chamber temperature setting
Example: Change the lock to ON from OFF (factory setting)
Description of operation Key operated Indication after operation
----- The current chamber temperature is
displayed.
1 Press scroll key for 5 seconds.
The current setting (L 0) is displayed
and the first digit blinks.
2 Press up arrow key once.
The display is changed to L 1 from
L 0.
3 Press set key. SET
Lock status is memorized and the
current chamber temperature is
displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.
17
DEFROST OF EVAPORATOR
The following 2 kinds of defrost methods are provided with this refrigerator. Both defrost methods are
controlled automatically.
Cycle defrost
To keep the chamber temperature stable, the refrigeration compressor is cycled on and off. During “off”
period any frost which has accumulated on the evaporator is melted by energizing a defrost heater. This
will not have any discernible effect on the chamber temperature.
Forced defrost
When the ambient humidity is high, or a large amount of damp product is being stored inside the chamber,
there is a possibility that cycle defrost may not be enough to remove all of the frost on the evaporator. In
this case, a forced defrost cycle can be initiated.
When the unit is operating under a forced defrost cycle, the current chamber temperature and dF are
displayed alternately on the temperature display.
Once the forced defrost cycle is completed, normal operation resumes.
CAUTION
The unit may collect excessive frost on the evaporator if it is installed in high temperature and high
humidity location. For example, the unit starts to defrost once a week with 2 oC setting in the ambient of
35 oC and 80 % R.H. The chamber temperature goes up to approximately 10 oC temporarily
during defrosting.
18
ALARM TEMPERATURE SETTING
Setting of high temperature alarm
By setting the high temperature alarm, the alarm indicator and the temperature display blink and the
alarm buzzer operates (after 15 minutes) when the chamber temperature is over the setting of high
temperature alarm. Set the high temperature alarm to protect the stored items against the damage
resulting from temperature rise.
Setting range of high temperature alarm:
Between chamber temperature plus 2 oC and chamber temperature plus 14 oC
Initial setting (factory setting): chamber temperature plus 5 oC
Example: Change the high temperature alarm to chamber temperature plus 3 oC from chamber
temperature plus 5 oC
Following shows a sample setting. Set the desired high temperature alarm according to the condition
of use.
Description of operation Key operated Indication after operation
1
----- The current chamber temperature is
displayed.
2 Press up arrow key for 5 seconds.
F00 is displayed and the first digit
blinks.
3 Press up arrow key once.
The display is changed to F01 from
F00.
4 Press set key. SET The current setting (005) is displayed
and the first digit blinks.
5 Press up arrow key and scroll the
figure to 3. The display is changed to 003 from
005.
6 Press set key. SET
Set temperature is memorized and
the current chamber temperature is
displayed.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function). In this case, the setting is not accepted.
<Important>
The high temperature alarm may be activated after defrosting or a large amount of load is stored in the
chamber. This is not a malfunction. The high temperature alarm is cancelled automatically when the
chamber temperature reaches the set temperature.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48

Phcbi MPR-1014R Kasutusjuhend

Tüüp
Kasutusjuhend
See juhend sobib ka